Displayed Average Noise Level (Danl); Procedure - Keysight N9010B Service Manual

Exa signal analyzer
Hide thumbs Also See for N9010B:
Table of Contents

Advertisement

Functional Tests

Displayed Average Noise Level (DANL)

Displayed Average Noise Level (DANL)
Test Limits (with 0 dB input attenuation)
See
The Displayed Average Noise Level (DANL) of the signal analyzer is measured
across a 10 kHz frequency span at several center frequencies. The analyzer
input is terminated into a 50 Ω load. A test is performed to assure the
measurement is not performed in the presence of a residual response. The
measurement is then averaged, and the result is normalized to a 1 Hz
bandwidth.
Table 17-2
DANL Measurement Requirements
Item
Ω
Termination, 50
Type-N (m)
Ω
Termination, 50
2.4 mm (f)
a. For Options 503, 507, 513, 526
b. For Options 532, 544
Figure 17-1
DANL Test Setup

Procedure

1. Configure the equipment as shown in
2. Press MODE/MEAS. Select Spectrum Analyzer mode, Swept SA
3. Set up the signal analyzer by pressing:
566
Table 17-2
for values.
Critical Specifications
(for this test)
Frequency: 10 MHz to 18 GHz
a
Frequency: 10 MHz to 50 GHz
b
measurement, Normal view. Press OK. Press Mode Preset on the analyzer.
FREQ, Center Frequency, 10, MHz
Input/Output, Input, RF Coupling, select DC
FREQ, Span, 10, kHz
AMPTD, –70, dBm
AMPTD, Attenuation, Mech Atten, 0, dB
Recommended
Keysight Model
909A Option 012
85138B
Figure 17-1
Keysight N9010B EXA Signal Analyzer Service Guide