Lost Can Partner - Agilent Technologies G1314B User Manual

Agilent 1260 infinity variable wavelength detector
Table of Contents

Advertisement

Probable cause
1
2
3

Lost CAN Partner

Error ID: 0071
During an analysis, the internal synchronization or communication
between one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one
or more of the modules is no longer recognized as being connected to the
system, the error message is generated.
Probable cause
1
2
3
Agilent 1260 Infinity VWD User Manual
Not-ready condition in one of the
instruments connected to the remote line.
Defective remote cable.
Defective components in the instrument
showing the not-ready condition.
CAN cable disconnected.
Defective CAN cable.
Defective main board in another module.
Error Information
General Error Messages
Suggested actions
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up
correctly for analysis.
Exchange the remote cable.
Check the instrument for defects (refer to the
instrument's documentation).
Suggested actions
• Ensure all the CAN cables are connected
correctly.
• Ensure all CAN cables are installed
correctly.
Exchange the CAN cable.
Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.