Error Information
Detector Error Messages

Flash Lamp Current Overflow

The lamp current of the xenon flash lamp is monitored constantly. If the
current gets too high, an error is generated and the lamp is turned OFF.
Probable cause
1
2

Flash Trigger Lost

This message is displayed when the flash trigger is no longer generated.
Probable cause
1
2
3
140
Shortage of trigger pack assembly or
defective FLL board.
Shortage of flash lamp assembly.
Firmware problem.
Multi Mode Off
Defective encoder.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Reboot the detector (power cycle).
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Agilent 1260 FLD User Manual