Storing the settings permanently with Bootp

If you want to change parameters of the detector using the Bootp follow the
instructions below.
1 Turn off the detector.
2 Change the detector's settings of the Configuration Switch to "Bootp &
3 Start the Agilent Bootp Service and open its window.
4 If required, modify the parameters for the detector according to your needs
5 Press OK to exit the Bootp Manager.
6 Now turn on the detector and view the Bootp Server window. After some
7 Close the Agilent Bootp Service and turn off the detector.
8 Change the settings of the detector's Configuration Switch to "Using
9 Power cycle the detector. The detector can be accessed now via LAN
1100 Series DAD and MWD User Manual
Store" mode, see
Table 6
using the existing configuration.
time the Agilent Bootp Service will display the request from the LAN
interface. The parameters are now stored permanently in the non-volatile
memory of the detector.
Stored" mode, see
Table 6
without the Agilent Bootp Service, refer to
Setup"
on page 54.
on page 35.
on page 35.
"PC and Agilent ChemStation
LAN Configuration