Forced Cold Start Settings

Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning switches 1 and 2 to other positions (other than being both
up) will allow for normal operation.
Forced cold start erases all methods and data stored in the non-volatile memory.
C A U T I O N
Exceptions are diagnose and repair log books which will not be erased.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
Table 46
Mode Select
TEST/BOOT
To return to normal operation, set switches back to your GPIB or RS 232
configuration settings.
264
Parity Settings
Parity
8
0
No Parity
0
Odd Parity
1
Even Parity
Forced Cold Start Settings
1
2
3
1
1
0
4
5
6
0
0
0
1100 Series DAD and MWD Reference Manual
7
8
0
1