Configuring The Msd Through The Web User Interface (Wui); To Change The Network Settings Of The Msd - Agilent Technologies 5977C Series Operating Manual

Table of Contents

Advertisement

4
Operating in EI Mode

Configuring the MSD through the Web User Interface (WUI)

Configuring the MSD through the Web User
Interface (WUI)
If your GC does not support LVDS communications with an Agilent GC, you can
use the WUI to configure the MSD network settings. Reasons that a GC does not
support configuring a 5977C MSD's network settings from the GC control panel
include any of the following:

To change the network settings of the MSD

This procedure assumes that the operator has access to a PC located on the
same LAN subnet as the MSD.
1 Open the top hinged cover on the MSD for accessing the analyzer to view the
2 Press the MSD On/Off switch to start the instrument. When the instrument
3 Copy the IP address, gateway, and subnet mask from the mini display
80
LVDS communication does not exist between the GC and MSD.
The GC is not an Agilent 8890 or 9000 model with the correct firmware.
eModule mini display readout.
has completed its startup initialization it displays the current IP address
information in the mini display readout and cycles through it for about
10 minutes.
readout. The defaults for each are 192.168.254.12, 0.0.0.0, and
255.255.255.0 respectively.
5977C GC/MSD Operating Manual