Using the Detector

Special Settings of the Detector

Special Settings of the Detector

In this chapter special settings of the detector are described.

Control Settings

Figure 24
Detector control settings
66
• Lamp: turn on and off of UV-lamp.
• At Power On: automatic lamp-on at
power on.
• Error Method: take error method or
current method (in case of an error).
• Analog Output Range: can be set to
either 100 mV or 1 V full scale, see
"Analog Output Settings"
• Lamp Type: can be set to either
G1314-60100 (standard VWD lamp) or
2140-0590 (DAD lamps), see also
"Exchanging a Lamp"
• Automatic Turn On: lamps can be
programmed (detector must be on for
this).
• Help: online help.
Agilent 1260 Infinity VWD User Manual
on page 69.
on page 123.