Error Information
Detector Error Messages
Illegal Temperature Value from Sensor on Main
Board
Error ID: 1071
This temperature sensor (located on the detector main board) delivered a value
outside the allowed range. The parameter of this event equals the measured
temperature in 1/100 centigrade. As a result the temperature control is switched
off.
Probable cause
1
2
Illegal Temperature Value from Sensor at Air Inlet
Error ID: 1072
This temperature sensor delivered a value outside the allowed range. The
parameter of this event equals the measured temperature in 1/100 centigrade.
As a result the temperature control is switched off.
Probable cause
1
2
Agilent InfinityLab LC Series VWD User Manual
Defective sensor or main board.
Detector is exposed to illegal ambient condi-
tions.
The temperature sensor is defect.
Detector is exposed to illegal ambient condi-
tions.
Suggested actions
Please contact your Agilent service representa-
tive.
Verify that the ambient conditions are within the
allowed range.
Suggested actions
•
Replace the cable to the main board.
•
Please contact your Agilent service represen-
tative.
Verify that the ambient conditions are within the
allowed range.