Calibration/Adjustment; Main Power Supply Voltages; Voc Supply Voltage - Siemens SIRECUST 1280 Service Manual

Bedside monitors
Table of Contents

Advertisement

E331 .E2501 .061.21.XX.02
System
SIRECUST
1280/1281
Bedside Monitor - Service Manual
7.
CALIBRATION/ADJUSTMENT
This
chapter
details
calibration
pro-
cedures
to
be
performed
anytime
a
printed
circuit
board
in
the
display
subsection is replaced, or if the monitor
fails functional
verification
testing.
The
procedures are used to test the display
subsection for proper operation, align the
display, and calibrate the monitor so that
it uses minimum power.
Calibration procedures require access to
potentiometers
located
on
the
printed
circuit boards within the monitor.
* Open and close the monitor only in
accordance with procedures outlined
in Chapter 4, Sections 4.1 and 4.2.
* Perform
the
appropriate
functional
verification
procedures
outlined
in
Chapter
5
before
returning
the
monitor to normal bedside service.
Boards comprising the display subsection
of the monitor are identified in Table 7-2.
Refer
to Table
7-3,
and
Figure
7-2
or
Figure 7-3 for identification and location .
of specific potentiometers
referenced
in
the procedures outlined in this chapter.
7.1.
Main Power Supply Voltages
NOTE:
Use monitor chassis as ground. Ad-
just power supply voltages with the monitor
configured for a normal operational load, i.e.
SIREMS
equipped
with
cartridges
and
all
Boards installed.
7.1.1.
+5 Voc Supply Voltage
1.
Connect a digital voltmeter (DVM)
to the
right leg
of R17
on
the
Display Processor Board. Refer to
Figure 7-1.
2.
Set R3 of the Power Supply Board
to +5 +0.05 Vi.
3.
Secure R3 with Torque-Seal® or
an
equivalent
substance
to
prevent any rotation and change
in the setting.
Table 7-1.
Recommended Tools and Test Equipment, Calibration/Adjustment
TOOL/EQUIPMENT
Current Shunt, Mono (for 1280)
Current Shunt, Color (for 1281)
12 inch Video Alignment Grid
VME
Extender Board
POLYMED/LIM Simulator
Waveform Simulator
Digital Multi-Meter (DMM), 47% digit
High Voltage Probe (for DMM), 1000:1
DESCRIPTION
SIEMENS Art. No. 96 60 606 EE999
SIEMENS Art. No. 96 60 614 EE999
SIEMENS Art. No. 84 28 450 RE999
SIEMENS Art. No. 87 91 568 E2524
SIEMENS Art. No. 45 28 345 EH436
Model
Dynatech-Nevada,
215A
(or
equivalent)
Fluke, Model 8050A (or equivalent)
Fluke 80K-40 (or equivalent)
Insulated Adjustment Wand, long handie w/ 1.8mm - 2.5mm wide blade
Bolt, M4 X 35