9 - 12
MO unit
•
Check that correct media is used, see spare part list.
•
Check that write protection is not activated.
All problems with MO disk
•
Identify MO disk, see Page 5 - 19.
If more than one disk type or disk number is reported please contact the headquarter
support center, HSC 24/USC, for investigation.
•
Check and, if necessary, repair MO disk, see Page 5 - 19.
•
Make sure that the SCSI cables are properly connected and fixed to the units and that
the cable is not bent to much just behind the MO unit, see Page 6 - 2.
•
Check that the MO unit is terminated correctly.
If the message "Failed to store on MO disk" appears
•
Make sure that local storage disk is inserted.
•
Make sure that MO unit is switched on.
•
Identify MO disk, see Page 5 - 19.
•
Check if MO disk is full.
•
Check and, if necessary, repair MO disk, see Page 5 - 19.
•
Try to save on a new MO disk.
•
Power off and reboot the Opdimar workstation.
•
Check cables, termination. Change MO unit.
If the message "Cannot read disk Prepare disk?" appears
WARNING
•
If it is a new disk, press OK (this will format the disk). Else, press Cancel.
•
Identify MO disk, see Page 5 - 19.
•
Check and, if necessary, repair MO disk, see Page 5 - 19.
•
Power off and reboot the Opdimar workstation.
•
Check cables and termination, see MAMMOMAT 1000/3000/3000 Nova - Opdimar
Maintenance Instructions.
•
Change MO unit.
Cannot communicate with MO unit
Test of SCSI bus
WARNING
Opdimar
Service
Prepare disk can not be performed on a new 2.3 GB MO-disk if
using a 9.1 MO unit.
Test of SCSI bus can destroy the hard disk content. The software
may have to be reinstalled.
Register 5
SPB7-230.061.11
Rev. 02
Troubleshooting guide
Page 12 of 14
07.03
CS PS 24
9
9
9
9
9