Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
Indicates a risk of danger that if disregarded leads or may lead to a potential situation which may result in an undesirable result or state (not death, physical injury or property damage). Page 5 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 6
Should be understood as a tip. The user does not absolutely have to observe these instructions. However, there will be advantages if he does. Page 6 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
Therefore before working on one of these boards, make sure that the LED V400 has gone out. This is the case 3 minutes after the ARCADIS Varic system is switched off. Page 8 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06...
NOTE Laser emissions! There is no direct hazard to the eye (blinking reflex). Nevertheless do not look directly into the laser beam. Page 9 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Observe the safety information in ARTD-002.731.03... When working with the laser light localizer, do not view the laser beams directly with optical instruments. Page 10 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
The values, including the measuring points, must be recorded as first measured values in the protective conductor resistance protocol. The measuring procedure and the measuring device used (designation and serial num- ber) must also be documented. Page 11 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 12
Separate the page with the protocol from these instructions and file it in the "Protocols" register in the system binder or log book. Fig. 1: Measuring circuit for measuring the protective conductor resistance for units that are Page 12 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP...
- even if the limit value of 0.2 ohms is not exceeded. System leakage current measurement information Observe the instructions in the "Safety Rules for Installation and Repair" (ARTD-002.731.17 ...). Page 13 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 14
However, please note the minimum resolution of the system leakage current measuring device and any additional manufacturer information restricting the use of the measuring device. Page 14 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP...
Page 15
15)), the system must be insulated during the measurement and must not be touched. Fig. 3: Measuring circuit for direct measurement of the system leakage current in compliance Page 15 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 16
DIN VDE 0751-1/2001-10, Fig. C5 for protection class I. Pos. 1 = System Pos. 2 = Application part type B (if available) Pos. 3 = Measurement setup (integrated into measuring device) Page 16 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
Page 17
This must also be documented. Document the measuring procedure (differential measurement or direct measurement) Page 17 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 18
(insulation damage, damage from moisture, defective interference suppressor, etc.) - even if the limit value of 2.5 mA is not exceeded. Page 18 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP...
Note on the log book The log book is located in the monitor cart behind the keyboard. It is accessible if the cover on the back of the monitor cart is removed. Page 20 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06...
Varic (refer to current and voltage values stated on the line voltage label, back of moni- tor cart). • The required line internal resistance is reached (refer to ARCADIS Varic Planning Guide). • The selectivity of the fuse protection is assured - according to the relevant national standards.
(the power plug must be disconnected) and the voltage output plug of the UPS must be discon- nected as well. Page 22 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP...
2General Start-up Measuring the local line voltage • Measure the voltage and frequency at the intended outlet for the ARCADIS Varic sys- tem. • Compare the measured line voltage with the line voltage label on the back of the moni- tor cart.
Page 24
General Start-up Fig. 4: Monitor cart line voltage Pos. 1 Power plug Pos. 2 T1 and T2 Page 24 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
"The emergency stop switch has been released." • Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm movement and emergency stop. Page 25 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
If the inner, square field of the indicator is white, the temperature has not been exceeded. Remove the temperature indicator. If the indicator is discolored (inner field black), proceed according to IQ docu- ment RXD0-000-038.01.xx.xx. Page 26 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP...
General Start-up Function checks Operation of the ARCADIS Varic, see operating instructions • Stand and C-arm system movements • Semi-transparent slot and iris diaphragm in fluoroscopy procedure. • Collimation of the semi-transparent slot and iris diaphragm using fluoroscopy proce- dure.
• Document the function of the C-arm movements, brakes and warning signals in the test protocol, "C-arm movement and emergency stop" . Page 28 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP...
Please clarify in advance with the customer whether the following system configura- tions are required. Service login Fig. 6: Service login • Click on <Options> <Customer service> <Local service>. Page 29 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 30
Specific system settings Fig. 7: Authentication_ • Enter the 6-character password and click OK (see system folder for password). Page 30 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
Specific system settings Customer address configuration Fig. 8: Configuration_ • Click on <Configuration>. Page 31 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 32
Specific system settings Fig. 9: Configuration • Click <Next>. Fig. 10: Configuration_Local Host • Click <Site Info>. Page 32 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
"serial no." or "station name". • Click <Save>. • Click <Home>. Main system configuration • Start the service program and log in as described under "Service Login". Page 33 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 34
Specific system settings Fig. 12: Main_system • Click <Main system>. Fig. 13: Main system_ • Select ARCADIS Varic and click <Next>. Page 34 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
Page 35
- Buzzer on: Buzzer sound always during radiation. The country-specific regulations must be observed for the max. fluoroscopy time and the radiation blockage. Factory setting: Page 35 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
A maximum tube current limit is required - as far as known - only in Denmark. • Start service program and log in as described under "Service Login". Page 36 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP...
Page 37
Specific system settings Fig. 15: Main_system • Click <Main system>. Page 37 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 38
Specific system settings Fig. 16: Main system • Select ARCADIS Varic and click <Next>. Fig. 17: Main System_Configuration_Current Limitation • Click <Current limitation>. • Make country-specific changes (only Denmark) and click <Save>. • Click <Home>. • Perform a system shutdown.
Open the <Options> - <End Session> menu and click the “Shut- down system” button. Immediately thereafter, press the Power off button on the monitor cart. The system shuts down and switches off automatically. Page 39 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Select "Backup & Restore" in the “Service Software Home Menu” • Place the backup CD (located in the monitor cart service compartment) in the CD ROM drive. Fig. 18: Backup & Restore_Command Page 40 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
Page 41
Specific system settings • Select the <Backup> command. Fig. 19: Backup & Restore_Command_Backup • Select the respective drive under “Drives”. - Save to CD ROM (standard save) Page 41 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 42
• Select “SW-Settings02” under Packages • Click <Go> and wait until the message “Ready” appears in the footer. Fig. 21: Backup & Restore_Command_Backup_SW-Settings02 ASPIA settings Page 42 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
Page 43
It is not possible to back up the database (patient images)! Store the patient images in an archive. Saving to CD ROM as a long-term archive is prohibited. Page 43 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Plug in the USB cable. • Print a test page after the automatic test. • Follow the instructions on the CD-ROM from the manufacturer in the event of problems. Page 44 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
"Measurement point 22" of column "First measured value". Mark empty fields with "n.a." as "not applicable". • In row "(1) Meas. circuit", column "First measured value", enter the text "VDE0751-1 Image C2". Page 45 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 46
Complete the "First measured value" column of the "Protective conductor resistance" protocol. • If you want to complete the protocol with additional remarks, enter them in the "Remarks" table in the protocol addendum. Page 46 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
Page 47
Monitor cart, one screw at the left side cover Monitor cart, one screw at the right handle Monitor cart, one screw at the left handle Page 47 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
"Measuring device, ser. no.", column "First measured value". • In "Device leakage current" protocol, row "(1) Meas. circuit", column "First measured value", enter the text "VDE0751-1 Image C5". Page 48 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP...
Page 49
Complete the "First measured value" column of the "Device leakage current" protocol. • If you want to complete the protocol with additional remarks, enter them in the "Remarks" table in the protocol addendum. Page 49 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Final Work Steps Completing the "ARCADIS Varic Installation" protocol The manufacturer of this product requires information and the legislator demands proofs that a product delivered free of defects from the factory continues to possess the required and certified quality properties on installation and start-up.
Protocols System leakage current System: ARCADIS Varic Material number: 08080017 Serial number: ................. Customer-specific identification ................. number: Tab. 9 System leakage current First measured value Repeat measurement values System leakage current (Highest measured value) [mA] Line voltage during the mea-...
DIN VDE 0751-1/2001-10, Fig. C5 for protection class II. Pos. 1 = System Pos. 2 = Application part (if applicable) Pos. 3 = Measurement setup (integrated into measuring device) Page 56 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
Page 57
DIN VDE 0751-1/2001-10, Fig. C6 for protection class II. Pos. 1 = System Pos. 2 = Application part (if applicable) Pos. 3 = Measurement setup (integrated into measuring device) Page 57 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Protocols Comments: Tab. 10 Date Comments Name Signature Page 58 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP Medical Solutions...
Protocols C-arm movement and emergency stop System: ARCADIS Varic Material number: 08080017 Serial number: ................. Customer-specific identification ................. number: Tab. 11 C-arm movement and emergency stop protocol Movement Name Date Initials Yes/No Releasing/locking the brakes Downward movement, position Downward movement, position The warning signals sound as described.
Protocols Network checklist System: ARCADIS Varic Material number: 08080017 Serial number: ................. Customer-specific identification ................. number: Network checklist (1+) For SAP order no.:_______________________ Will the system have a network connection (HIS/RIS, Camera, PACS) ? Mandatory for image system with network connection! To ensure that the system is preconfigured at the factory and thus optimize installation at the customer site, we request that you complete and return this list if a network connection is available (and used).
Page 61
Logical name * Configuration/ DICOM/ Network nodes Application entity: AE title * Configuration/ DICOM/ Network nodes Port number * Configuration/ DICOM/ Network nodes Supported DICOM services: Page 61 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 63
General node properties: Logical name * Configuration/ DICOM/ Print devices Application entity: AE title * Configuration/ DICOM/ Print devices Port number * Configuration/DICOM/Print devices Remark: Page 63 of 64 Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic Medical Solutions 12.06 CS PS SP...
Page 64
The specification for the buzzer setting for the USA was changed as follows: Warning time limit: changed from "45" to "50" Blocking time limit: changed from "5" to "0" Page 64 of 64 ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG 12.06 CS PS SP...