Chapter 7 Maintaining the Instrument
7-48
Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide
September 1, 2004 11:39 am, CH_Maintenance.fm
9. Click
The software records the rotary position for the Zymark position 5 (input
stack 2).
10. Repeat
steps 1 through 9
for the remaining positions 6 and 7:
11. Exit the Zymark Twister Software.
a. Click
b. Click Exit.
The software closes.
A bug in the Zymark Twister Software can cause portions of the program to
persist in memory even after the software has been closed. Because the Zymark
Twister Software conflicts with the SDS software, these residual elements must
be closed inside the Windows Task Manager before continuing.
12. Confirm that the Zymark Twister Software has closed by viewing the Task
Manager.
a. Press the Ctrl + Alt + Del keys in unison.
b. In the Windows Security dialog box, click Task Manager.
c. In the Task Manager dialog box, confirm that the Twister software has
closed by looking for the Twister software entry in the Task list. If the
software is still running, click the software entry and click End Task to
exit the software.
d. Select File > Exit to quit the Task Manager.
13. Replace the covers for the fixed-position bar code reader and the underlying
platform (removed in
GR2009
7900HT
Front view with Robot
7900HT FAST Real-Time PCR System
GR2009
Fixed-position bar code reader
and underlying platform covers
DRAFT
, and click
.
for input stacks 3 and 4 to define Rotary Offset values
Zymark position 7
(input stack 4)
Zymark position 6
(input stack 3)
.
step 1 on page
7-41).