Chapter 7 Maintaining the Instrument
7-40
Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide
September 1, 2004 11:39 am, CH_Maintenance.fm
10. Exit the Zymark Twister Software.
a. Click
b. Click Quit Application to close the software.
A defect in the Zymark Twister Software can cause portions of the program to persist
in memory even after the software has been closed. Because the Zymark Twister
Software conflicts with the SDS software, the residual elements of the software must
be closed inside the Windows Task Manager before continuing.
1. Confirm that the stack has closed by viewing the Task Manager.
a. Press the Crtl + Alt + Del keys in unison.
b. In the
Windows Security dialog box,
c. In the Task Manager dialog box, confirm that the software has closed by
looking for the Zymark Twister Software entry in the Task list. If the
software is still running, click the software entry and click Close to exit the
remaining software.
d. Select File > Exit.
DRAFT
.
click Task Manager.