Applied Biosystems 7900HT User Manual page 263

Fast real-time pcr system and sds enterprise database
Hide thumbs Also See for 7900HT:
Table of Contents

Advertisement

Purpose of
TGF-β Runs
Materials
Required
Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide
Table 7-2
Configurations of the TaqMan RNase P Instrument Verification Plates
TaqMan RNase P
Instrument Verification
Plate
TaqMan
RNase P
®
Instrument Verification
Plate
(PN 4310982, standard
96-well plate)
The TaqMan
®
Low Density 7900HT Installation Array (TGF-β card) is an
experiment run to verify the performance of the 7900HT instrument for use with
TaqMan Low Density Arrays. The TGF-β card is preloaded with the reagents
necessary for the detection and quantification of genomic copies of the TGF-β gene.
Each well contains preloaded TGF-β primers and FAM MGB dye-labeled probe.
RNase P Runs
• Appropriate TaqMan RNase P Instrument Verification Plate (see
page
7-30)
• Centrifuge, with plate adaptor
TGF-β Runs
• TGF-β card
• TaqMan
®
2✕ Universal PCR Master Mix
• Water, DNase/RNase-free, deionized
• 2-mL microtube, sterile
• Centrifuge
• Microfluidic card sealer
Sample Configuration
Unknown 1
NTC
STD 1250
STD 5000
STD 10000
Unknown 2
DRAFT
September 1, 2004 11:39 am, CH_Maintenance.fm
Verifying Instrument Performance
5000
STD 2500
STD 20000
10000
Table 7-2 on
7-31

Hide quick links: