Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide
4. Apply pure dyes to the custom plate document:
a. Select the wells containing the custom dye.
b. In the Dyes drop-down list in the Setup tab, select the appropriate dye.
The software applies the dye to the selected wells.
c. Repeat steps a and b to configure the plate document with any additional
custom dyes.
5. Save the custom Pure Dye plate document as a plate document template:
a. Click
(or select File > Save).
b. In the Save dialog box, navigate to Program Files > Applied
Biosystems > SDS 2.2.1 > Templates.
The Templates directory appears inside the Look in field. By saving the
plate document template to the Templates directory, it becomes available
from the Template drop-down list in the New Document dialog box.
c. Select Files of type > SDS 7900HT Template Document (*.sdt).
d. In the File name field, and enter a name for the plate document template.
e. Click
. The software saves the plate document as a plate document
template.
6. Run the custom Pure Dye plate as explained in
on page
7-22.
Adding Custom Dyes to the Pure Dye Set
Dyes drop-down list
Custom dye added to selected
wells of the plate document
DRAFT
September 1, 2004 11:39 am, CH_Maintenance.fm
"Preparing the Pure Dye Plates"
7-29