:SBUS<n>:SPI:TRIGger:TYPE
Command Syntax
:SBUS<n>:SPI:TRIGger:TYPE <value>
<value> ::= {MOSI | MISO}
The :SBUS<n>:SPI:TRIGger:TYPE command specifies whether the SPI
trigger will be on the MOSI data or the MISO data.
When triggering on MOSI data, the data value is specified by the
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA and
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh commands.
When triggering on MISO data, the data value is specified by the
:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA and
:SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh commands.
Query Syntax
:SBUS<n>:SPI:TRIGger:TYPE?
The :SBUS<n>:SPI:TRIGger:TYPE? query returns the current SPI trigger
type setting.
Return Format
<value><NL>
<value> ::= {MOSI | MISO}
See Also