SBR Diagnostics

Submenu 1
Submenu 2
SBR Diagnostics
Length Calibration
Re-zero Thickness
Last 256 Dimensions
Reset SBR Status
Length Calibration
This program allows you to calibrate the length sensors (FS2/WS1, WS2,
WS3) in the feeder/WOW module. A length calibration gauge is required for
this test (PN# White DJ62084 or Black DW62384).
At the Calibration screen, place the length calibration gauge on feed deck
and adjust the side guide appropriately. Press Start to continue. Run one
calibration gauge through the machine, 16 successive times. You must do
each run individually, and it must be with the same gauge – if you load a
stack of 16 calibration gauges, the machine may give length or width errors
during this process. NOTE: If an error occurs or if calibration fails, check
the side guide setting to make sure the length calibration gauge is not being
skewed when feeding, then press OK to reattempt the length calibration.
Re-zero Thickness
This allows you to re-zero the thickness encoder (T1) in the WOW. You
must make sure nothing is anywhere on the mail deck during this proce-
dure. NOTE: It is normal for the encoder not to return fully to "0.00".
Submenu 3
Rest Status
Don't Reset status
DM300c/DM400c/DM475 Service Manual (SV61801 Rev. D)