Auxiliary Input orrection, Adjustment #111
his procedure shows you how to generate correction constants for the auxiliary input. he
procedure measures two voltages (0 and +10.000 volts) applied to the auxiliary input to
compute the o set and gain of the auxiliary input circuitry.
equired Equipment
Equipment
escription
Voltage reference source (part of service kit)
50 BNC cable
Note
he voltage reference source provides +10 V at its BNC connector. his
connector must be connected to the AUX input on the rear panel of the
analyzer.
Warmup Time
Allow the analyzer to warm up for one hour.
stimated Adjustment Time
his adjustment takes approximately one minute to perform.

Procedure

1. Press
PRESET
SYSTEM OPTIONS
5
Execute Test
111
ENTER
5
5
2. his is a fully automated adjustment handled by the internal rmware. Follow the
instructions on the analyzer's display.
not
3. If you are
going to make any more adjustments, press
Update Corr Const
are
If you
going to make more adjustments, save the correction constant data that you have
generated so far.
a. Insert a formatted disk into the internal disk drive.
b. Press
SYSTEM OPTIONS
a le (or writes over an existing le) where the correction constants are stored.
ervice Test and Adjustments
5
.
.
tore CC to EPROM
ervice Update Corr Const
5
Recommended Model or Part Number
08712-60031 or any +10.000 V voltage source.
any
SYSTEM OPTIONS
tore CC to Disk
elect Adjustment
ervice
5
. his creates
Adjustments
2-21