Error
Noisy threshold plot
Possible causes
Debris in sample line
Debris in sheath or DI water
bottles
Dead cells in sample
Acquisition/Event Issues
Troubleshooting steps
Follow these steps, testing
after each to determine
whether the problem has
been resolved:
1. Run the cleaning process.
See
Cleaning the Sample
Line and Probe on page
324.
2. Run the Decontamination
wizard. See
Decontaminating the
System on page
325.
Follow these steps, testing
after each to determine
whether the problem has
been resolved:
1. Run the cleaning process.
See
Cleaning the Sample
Line and Probe on page
324.
2. Run the Decontamination
wizard. See
Decontaminating the
System on page
325.
Adjust gates (see
Applying
Filters (Gates) on page
236)
or repeat experiment with
fresh cells.
User Guide | 355
Need help?
Do you have a question about the ZE5 and is the answer not in the manual?