Error
High event rate
High voltage present in certain
channels during QC process
Possible causes
Trigger channel voltage too high
Incorrect beads used
Acquisition/Event Issues
Troubleshooting steps
Decrease the trigger channel
voltage until data start to
appear in the threshold plot.
Then, adjust the trigger
voltage and threshold value
until data appear as expected
in the trigger channel plot in
the workspace. See
Configuring Instrument
Settings on page
257.
Ensure single peak
QC Beads are being used.
ProLine™ Rainbow Beads or
ProLine™ Universal
Calibration Beads do not
fluoresce in certain channels,
especially for the UV and
violet lasers.