Chapter 5 Configuring the System

Working with Optical Filter Configurations

To view and edit the optical filter configuration, click Display Optical Filter Configuration in the
Acquisition toolbar.
The layout of the detection path is shown on the left, reflecting the current configuration of the
instrument. If you install a different physical filter, you must edit the corresponding mirror/filter position
labels in this section of Everest Software (see
Editing the Optical Filter Configuration on page
132).
A green check mark on each final position in each optical detection path indicates that the ZE5-EYE
process has verified that the filter listed in Everest Software is currently installed in the instrument. The
ZE5-EYE relies on the filter labels used in the software. If you change a filter label without changing
120 | ZE5 Cell Analyzer and Everest Software