Safe Use Specifications and Compliance
Laser Product Hazard Classification
The intent of the laser hazard classification is to identify hazards to users posed by the laser and to
provide appropriate protective measures. The ZE5 Cell Analyzer is a Class 1 laser product that
complies with 21 CFR 1040.10 and 1040.11, except for deviations pursuant to Laser Notice No. 50,
dated June 24, 2007 stating that operators are not exposed to harmful levels of laser radiation during
normal operation, maintenance, and/or service. During times of repair and/or major service by a
trained technician, laser safety controls for Class 3B lasers must be followed.
WARNING! Use of controls, adjustments, or procedures other than those specified
herein may result in hazardous laser radiation exposure.
Electrical Safety Information and Classification
The ZE5 Cell Analyzer conforms to international regulations encompassing the accessibility of high
voltages by the user (IEC61010-1). Use all protective housings and shields as specified in this guide.
Further information about specific electrical hazards is listed in the hardware description.
AC Fuse Requirements
Remove power cord before replacing fuses. Fuses are 5 x 20 mm and must be rated to AC250V, 4 A
slow blow such as Schurter 0034.3123.
AC Power Cord Requirements
Power cord must be IEC 60320-1 compliant with a C13 plug on the instrument end. If the power cord
must be replaced, replace it with only an adequately rated cord.
Position the instrument for easy access to the power switch and the power cord.
AC Outlet Requirements
Three outlets (instrument, monitor, computer), AC100–240V, 50–60 Hz, <500 W.
Safe Use Specifications and Compliance
User Guide | 15
Need help?
Do you have a question about the ZE5 and is the answer not in the manual?