Chapter 8 Acquiring Samples
2.
To review the sampling configuration, click View in the Instrument Control panel.
The run list opens, summarizing the experiment setup. Click the X in the upper right corner to
close it.
3.
Click (Start) Run List to begin sampling.
The instrument acquires samples as determined in the Experiment Builder. If multiple limits are
set, acquisition stops when the first limit is reached.
After a position has been sampled, its position on the map is indicated by a black check mark. For
information about symbols shown in the plate map, see
on page
205.
Tip: By default, sampling begins at position A1 in the plate map, but you can begin
acquisition from any sample position. Click a position in the plate map and click (Start) Run
List. The run list proceeds as determined in the Experiment Builder, starting from the selected
position.
4.
To stop the run list before acquisition is complete for all positions, click (Stop) Run List.
284 | ZE5 Cell Analyzer and Everest Software
Plate Map in the Instrument Control Panel