Appendix C
Ports
Ports for connecting the ZE5™ Cell Analyzer to house DI water and house waste are located on the
instrument back panel. Contact Bio-Rad Technical Support if you plan to use either of these ports.
Before putting the instrument into external DI water mode or external waste mode, a qualified, trained
technician must hook up your house DI water or waste line to the instrument.
Caution: Only Bio-Rad Service personnel are permitted to connect the ZE5 Cell
Analyzer to external DI water or waste lines.

Connecting External DI Water

When the instrument is connected to external DI water and is put into external DI water mode, the
volume in the top DI water bottle will drop to around 1/4 full while sheath fluid is flowing, then the bottle
will automatically refill to around 3/4 full. The bottom DI bottle is not used in this mode. For information
about external DI water mode, see
Optional External DI Water and Waste
Specifying External Port Use on page
381.
User Guide | 379