A message in the Instrument Control Panel indicates that the system is paused.
To resume the system
u
In the System section on the Acquisition workspace toolbar, click Resume System.
Shutting Down
Important:
Do not shut down the ZE5 Cell Analyzer using the main power switch. Refer to
Power on page 141
Shut down the ZE5 Cell Analyzer system at the end of each day of use. This turns off all lasers; cleans
the sample line, probe, and flow cell with cleaner; depressurizes the fluidics; and logs you out of the
system. The shutdown process is entirely automated and does not require user interaction.
Important:
Always follow the personal protective equipment (PPE) guidelines relevant to your
laboratory's safety procedures for dealing with ethanol or bleach.
The Shutdown button is located
n
In the Home window
n
In the System Actions section on the Acquisition workspace toolbar
It can also be found in the Login window, if your system administrator has allowed shutdown to occur
when no user is logged in. For more information, see
for more information.
Specifying Logged Out Settings on page
Shutting Down
System
114.
User Guide | 161