Chapter 12 Maintenance
Cleaning the Sample Line and Probe
Everest Software includes an option to initiate a sample line and probe clean cycle. This process runs
the onboard system cleaner through the sample line and probe. You can also perform the cleaning
using a tube of ethanol or bleach solution. See
cleaning solutions to use in the sample line.
To clean the sample line and probe
1.
In the System section of the toolbar, click Clean the Sample Probe in the Instrument Tools
dropdown list..
2.
Follow the directions that appear on screen.
The system is ready for use when this process is complete, but you can choose to run the QC
process afterward.
324 | ZE5 Cell Analyzer and Everest Software
Cleaning Solutions on page 321
for recommended