Determine whether the probe has entered the first position on the custom device.
n
If the probe is centered in the first position, click Yes.
n
If the probe is not centered in the first position, click No, use the arrows to adjust it to the
center of the well, and click Accept Current Location. Repeat this process as needed.
9.
Determine whether the probe has entered the last position on the custom device.
n
If the probe is centered in the last position, click Yes.
n
If the probe is not centered in the last position, click No, use the arrows to adjust it to the
center of the well, and click Accept Current Location. Repeat this process as needed.
The layout of the custom device appears in the Plate Setup panel.
To select a custom plate
1.
In the Media list, click Custom.
2.
Select the custom plate from the list of plates and then click Select.