GPIB Interface

GPIB Interface

Cables and Adapters

Keysight N774x Series
is no DHCP server, for example when the instrument is connected
directly to a PC.
The LAN Reset button
This recessed button has two functions.
Pressing the button briefly invokes a preset of the instrument and
restores default measurement settings. This is equivalent to the
programming command
:SYSTem:PRESet
Pressing and holding the button for 3 seconds will reset the LAN
parameters to the factory default. This includes changing the
password for modifying the configuration via LAN back to the
default "Keysight". This is equivalent to the following sequence of
programming commands:
:SYSTem:COMMunicate:ETHernet:RESet
You can connect your GPIB interface into a star network, a linear
network, or a combination star and linear network. The limitations
imposed on this network are as follows:
The total cable length cannot exceed 20 meters.
The maximum cable length between devices is 4 meters, with an
average separation of 2m between devices over the whole bus.
No more than 15 devices may be interconnected on one bus.
For further details on cables and adapters, refer to "Cables" section in
the User's Guide.
1
Getting Started
27