Installing the Detector

Flow Connections to the Detector

Flow Connections to the Detector

Tools required
Description
Wrench, 1/4 – 5/16 inch
(for capillary connections)
Parts required
#
1
Hardware required
Other modules depend on system setup
Preparations
Detector is installed in the LC system.
Toxic, flammable and hazardous solvents, samples and reagents
WA R N I N G
The handling of solvents, samples and reagents can hold health and safety risks.
The flow cell is shipped with a filling of isopropanol (also recommended when the
N O T E
instrument and/or flow cell is shipped to another location). This is to avoid breakage due to
subambient conditions.
52
p/n
G1314-68755
When working with these substances observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in
the material handling and safety data sheet supplied by the vendor, and follow good
laboratory practice.
The volume of substances should be reduced to the minimum required for the
analysis.
Do not operate the instrument in an explosive atmosphere.
Description
Accessory kit
Agilent 1200 Infinity Series VWD - User Manual