Introduction

Detector Setup

Front Side
Figure 3 Front view with open front cover
The following connections and indication elements are accessible on the front
panel:
1 Power On LED
2 Status LED
3 Inlet Port
4 Outlet Port
Never block the outlet port if any flow is attached to the inlet port. This can lead to a
CAUTION
high pressure inside the detector which causes internal leakages.
Multi-Angle Light Scattering Detector User Manual
If the system is powered on and running, the small LED is illuminated.
The two LEDs show the actual status of the device:
LASER ON: green illuminated if the laser inside the detector is active.
ERROR: red illuminated if an internal vapor or leakage error occurred, or an
external error is notified.
The error state and the status of the laser will also be indicated via the software.
This port is the flow inlet of the detector.
This port is the flow outlet of the device which can be connected to the next
detector or to the waste bottle.