Replacing The Detector's Firmware - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector

Replacing the Detector's Firmware

The installation of new firmware is required:
• if new version solves problems of currently installed version, or
• if after exchange of the detector main board (DAM or MWM) the version on
To upgrade the detector's firmware the following steps have to be performed:
1 Load the firmware into the detector, see help system of your user interface.
The G1315B DAD and the G1365B MWD require firmware revision 4.x or above (main and
N O T E
resident).
2 Perform a
The wavelength recalibration should be performed after at least 10 minutes of warm-up
N O T E
time of the lamp.
3 If detector main board was replaced, reenter the serial number information
180
board is older than previous installed one.
"Wavelength Recalibration"
parameters into the board's memory.
of the module through the user interface, see
using the Control Module"
using the Agilent ChemStation"
on page 84 to add the recalibration
"Entering the Serial Number
on page 146 or
"Entering the Serial Number
on page 147.
1100 Series DAD and MWD Reference Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents