Grating And Diode Array - Agilent Technologies InfinityLab LC Series User Manual

Diode array detectors
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

1
Introduction
Optical System

Grating and Diode Array

The combination of dispersion and spectral imaging is accomplished by using a
concave holographic grating. The grating separates the light beam into all its
component wavelengths and reflects the light onto the photodiode array.
The diode array is a series of 1024 individual photodiodes and control circuits
located on a ceramic carrier. It has a wavelength range from 190 – 640 nm and
the sampling interval is 0.5 nm.
Figure 9
Agilent InfinityLab LC Series Diode Array Detectors User Manual
Grating and diode array
18