Replacing Capillaries On A Standard Flow Cell - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector

Replacing Capillaries on a Standard Flow Cell

When required
Tools required
Parts required
Preparations for this procedure:
Turn off the flow.
Remove the front cover.
Remove the flow cell, see
page 105.
114
If the capillary is blocked
Two 1/4 inch wrenches for capillary connections
Wrench 4 mm for capillary connections
Screwdriver Pozi 1 PT3
For parts see
"Exchanging a Flow Cell"
"Standard Flow Cell"
on page 186
1 Identify the inlet and outlet capillaries.
Inlet
on
capillary
1100 Series DAD and MWD Reference Manual