4
Repairing the Detector

Exchanging the Fan (G1315A/G1365A)

Use this procedure for the the G1315A/G1365A only. For G1315B/G1365B detectors refer
N O T E
to
When required
Tools required
Parts required
Preparations for this procedure:
Turn off the lamp.
Switch off the detector, and disconnect the cables and
acpillaries.
Remove detector from stack and place it on the working
bench.
Remove front cover, top cover and top foam section, see
"Removing the Top Cover and Foam"
148
"Replacing Fan Parts (G1315B/G1365B)"
If the fan is defective or noisy
Screwdriver POZI 1 PT3
Fan assembly 3160-1016
on page 140.
on page 150.
1 Disconnect the connectors from the main board (fan J16)
and pull the fan assembly out of its location.