Installing The Control Module; Figure 55 Y-Rail In Unlocked, Functional Position - Quantum scalar i500 User Manual

Hide thumbs Also See for scalar i500:
Table of Contents

Advertisement

Figure 55 Y-Rail in Unlocked,
Functional Position
Installing the Control
Module
Scalar i500 User's Guide
7 Repeat these steps for each expansion module you are installing.
Install the 5U control module as follows:
12
1 Open the control module's I/E station door and access door.
2 Lift the control module and place it in the desired location.
3 If stacking the control module on top of another module, secure the
two modules together by tightening the two thumbscrews at the base
of the front of the module and the two thumbscrews located at the
base of the back of the module. Then lower the module's guide pin
(located at the base of the front of the module) by turning it and
pushing it down.
4 Tighten all thumbscrews located at the base of the front and back of
the modules.
5 Use the rack ears to fasten the control module to the rack.
Chapter 12 Installing, Removing, and Replacing
Installing a New Multi-Module Library Configuration