Trigger
Remote Command
Notes
Preset
State Saved
Min
Max
Default Unit
Initial S/W Revision
Sync Holdoff
Sync Holdoff specifies the duration that the sync source signal must be kept false before the transition to
true to be recognized as the sync timing. The periodic timer phase is aligned when the sync source signal
becomes true, after the Holdoff time is satisfied.
A holdoff of 2 ms will work with most WiMAX signals, but there may be cases where the burst off
duration is less than 1 ms and this value will need to be changed.
Key Path
Remote Command
Preset
State Saved
Min
Max
Default Unit
Initial S/W Revision

Baseband I/Q

Pressing this key when it is not selected selects Baseband I/Q as the trigger. Pressing the key when it is
already selected accesses the Baseband I/Q trigger type selection menu. The key is annotated to display
which of the Baseband I/Q trigger types is currently selected.
598
:TRIGger[:SEQuence]:FRAMe:DELay <time>
:TRIGger[:SEQuence]:FRAMe:DELay?
:TRIGger[:SEQuence]:FRAMe:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:FRAMe:DELay:STATe?
Note that delay is used when the sync source is not set to OFF. If the sync
source is set to OFF, offset is used.
Off, 1.000 us
Saved in instrument state
–150 ms
+500 ms
s
Prior to A.02.00
Trigger, Periodic Timer
:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff <time>
:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff?
:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:FRAMe:SYNC:HOLDoff:STATe?
On, 1.000 ms
Saved in instrument state
0 ms
+500 ms
s
Prior to A.02.00