Remote Command
Example
Dependencies
Preset
State Saved
Min
Max
Initial S/W Revision
Gate Delay
Controls the length of time from the time the gate condition goes True until the gate is turned on.
Key Path
Remote Command
Example
Notes
Preset
State Saved
Min
[:SENSe]:SWEep:EGATe:TIME <time>
[:SENSe]:SWEep:EGATe:TIME?
SWE:EGAT:TIME 500 ms
Gate View Sweep Time is initialized:
On Preset (after initializing delay and length).
Every time the Gate Method is set/changed.
Additionally, in the Swept SA measurement, whenever you do a Preset, or
leave Gate View, the analyzer remembers the Gate Delay and Gate Length
settings. Then, when returning to Gate View, if the current Gate Delay and/or
Gate Length do not match the remembered values Gate View Sweep Time is
re-initialized.
1. Compute the location of the "gate stop" line, which you know is at time t =
t
+ GateDelay + GateLength.
min
519.3 µs
WiMAX OFDMA: 5 ms
GSM/EDGE: 1 ms
Saved in instrument state
1 µs
6000 s
Prior to A.02.00
Sweep/Control, Gate
[:SENSe]:SWEep:EGATe:DELay <time>
[:SENSe]:SWEep:EGATe:DELay?
SWE:EGAT:DELay 500ms
SWE:EGAT:DELay?
Units of time are required or no units; otherwise an invalid suffix error
message will be generated. See error –131.
57.7 us
WiMAX OFDMA: 71 us
GSM/EDGE: 600 us
Saved in instrument state
0.0 us
Sweep/Control