Sets the skew factor for the I channel. The skew will shift the channel's data in time. Use this to
compensate for differences in the electrical lengths of the input paths due to cabling.
Key Path
Remote Command
Example
Preset
State Saved
Range
Initial S/W Revision
I Probe
Access the probe setup parameters for the I channel. See
Key Path
State Saved
Readback Text
Initial S/W Revision
420
Input impedance is a built-in characteristic of a probe. Therefore, whenever a
probe is sensed, this key is disabled and the value is set to match the probe.
When no probe is sensed on Q and Q Same as I is On, the value set for I will
also be copied to Q.
LOW
Yes
This is unaffected by a Preset but is set to the default value on a "Restore
Input/Output Defaults" or "Restore System Defaults->All"
50 Ω | 1 MΩ
Prior to A.02.00
Input/Output, I/Q, I Setup
[:SENSe]:CORRection:IQ[:I]:SKEW <seconds>
[:SENSe]:CORRection:IQ[:I]:SKEW?
Delay the data for the I channel by 10 ns.
CORR:IQ:SKEW 10 ns
0
Yes
This is unaffected by Preset but is set to the default value on a "Restore
Input/Output Defaults" or "Restore System Defaults->All"
0 s to 100 ns
Prior to A.02.00
"I/Q Probe Setup" on page
Input/Output, I/Q, I Setup
No
[<I port probe id>]
This is reporting the type of probe sensed on the I port. There is no parameter
for overriding what is sensed.
Prior to A.02.00
424.