Agilent Technologies X Series Manual page 606

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Trigger
State Saved
Initial S/W Revision
Trig Delay
Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger
criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT.
Key Path
Remote Command
Example
Preset
State Saved
Range
Initial S/W Revision
Auxiliary Channel I/Q Mag
Pressing this key, when it is not selected, selects the Auxiliary Channel I/Q magnitude signal as the
trigger. The Auxiliary Channel I/Q Magnitude trigger condition is met when the auxiliary receiver's I/Q
magnitude output crosses the Auxiliary I/Q magnitude trigger level.
Key Path
Example
Readback Text
Initial S/W Revision
Trigger Level
Sets a level for the I/Q magnitude trigger. When the signal crosses this level, with the chosen slope, the
trigger occurs.
Key Path
Remote Command
606
Saved in instrument state
Prior to A.02.00
Trigger, Baseband I/Q, Input Q
:TRIGger[:SEQuence]:QINPut:DELay <time>
:TRIGger[:SEQuence]:QINPut:DELay?
:TRIGger[:SEQuence]:QINPut:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:QINPut:DELay:STATe?
TRIG:QINP:DEL 10 ms
TRIG:QINP:DEL:STAT ON
1 us
OFF
Saved in instrument state
–2.5 s to +10 s
Prior to A.02.00
Trigger, Baseband I/Q
TRIG:<meas>:SOUR AIQM
Aux I/Q Mag
Prior to A.02.00
Trigger, Baseband I/Q, Aux Channel I/Q Mag
:TRIGger[:SEQuence]:AIQMag:LEVel <ampl >
:TRIGger[:SEQuence]:AIQMag:LEVel?