After saving your changes, the analyzer asks you if you would like it to restart so that your changes can
take effect (see dialog box, below). If you choose not to restart, no memory will be released until the next
time you shut down and restart the analyzer.
Key Path
Remote Command
Example
Notes
Notes
Initial S/W Revision
Modified at S/W Revision
Exit Without Saving
Pressing this key will exit the Configure Applications utility without saving your changes.
Key Path
Initial S/W Revision
Modified at S/W Revision
Configure Applications - Instrument boot-up
At start-up of the analyzer program dialog box similar to the one under the
Configure Applications
to be loaded. This dialog will only be displayed if the memory required to pre-load all of the licensed
applications exceeds the Virtual Memory available.
System, Power On, Configure Applications
:SYSTem:PUP:PROCess
:SYST:PUP:PROC
You must Wait after this command for the instrument application to restart
The softkey will be grayed-out when the virtual memory of the selected
applications exceeds 100% of the limit.
You cannot use *WAI or *OPC? to synchronize operation after a restart. This
command stops and restarts the instrument application, thus the SCPI
operation is terminated and restarted. A remote program must use fixed wait
time to resume sending commands to the instrument. The wait time will be
dependent upon which applications are pre-loaded.
A.02.00
A.04.00
System, Power On, Configure Applications
A.02.00
A.04.00
key will be displayed allowing you to choose which licensed applications are
This is the SCPI command for restarting the analyzer.
System, Power On,
System Functions