Initial S/W Revision
Next
Perform the Q port calibration.
Key Path
Remote Command
Example
Notes
Notes
State Saved
Initial S/W Revision
Exit
Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration
data.
Key Path
Notes
Initial S/W Revision
Q-bar Port
The Q-bar port calibration is performed with the front panel's Q-bar port connected via a short BNC
cable to the Cal Out port. The guided calibration will show a diagram of the required connections.
Back
Return to the prior step in the calibration procedure.
Key Path
Notes
Prior to A.02.00
Input/Output, I/Q, I/Q Cable Calibrate...
:CALibration:IQ:FLATness:Q
CAL:IQ:FLAT:Q
The recommended procedure is to use the same BNC cable to calibrate all I/Q
ports. All I/Q ports should be calibrated sequentially during the procedure.
The calibration data is saved as soon as the port is calibrated and will survive
power cycles. It is not reset by any preset or restore data commands.
The Q port must be connected to the Cal Out port before issuing the SCPI
command.
No
Prior to A.02.00
Input/Output, I/Q, I/Q Cable Calibrate...
Using the Exit button will not restore the calibration data to the state prior to
entering the guided calibration. Once a port is calibrated the data is stored
immediately and the only way to change it is to redo the calibration step.
When the calibration may be left in an inconsistent state, a confirmation
dialog will be displayed (see
Prior to A.02.00
Input/Output, I/Q, I/Q Cable Calibrate...
Using the Back button will not restore the calibration data to a prior state.
Once a port is calibrated the data is stored immediately and the only way to
change it is to redo the calibration step. The Back button allows the user to go
back to a prior step to redo that calibration step.
"Exit Confirmation" on page
Input/Output
482).