Agilent Technologies X Series Manual page 422

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Input/Output
Q Same as I
Many, but not all, usages require the I and Q channels have an identical setup. To simplify channel setup,
the Q Same as I will cause the Q channel parameters to be mirrored from the I channel. That way your
only need to set up one channel (the I channel). The I channel values are copied to the Q channel, so at
the time Q Same as I is turned off the I and Q channel setups will be identical. This does not apply to
Probe settings or to parameters that determined by the probe.
Key Path
Remote Command
Example
Couplings
Preset
State Saved
Range
Readback Text
Initial S/W Revision
Q Differential Input
Selects differential input on or off for the Q channel. For differential input (also called balanced input),
the analyzer uses both the Q and Q-bar ports. When differential input is off (also called single-ended or
unbalanced input), the analyzer uses only the Q port.
Key Path
Remote Command
Example
422
Input/Output, I/Q, Q Setup
:INPut:IQ:MIRRored OFF|ON|0|1
:INPut:IQ:MIRRored?
Turn off the mirroring of parameters from I to Q.
INP:IQ:MIRR OFF
Only displayed for the Q channel. When Yes, the I channel values for some
parameters are mirrored (copied) to the Q channel. However, when a
parameter is determined by the type of probe and a probe is sensed, the probe
setting is always used and the I channel setting is ignored. The following
parameters are mirrored:
Differential Input (when not determined by probe)
Input Z (when not determined by probe)
This is unaffected by a Preset but is set to the default value (Q Same as I set to
"On") on a "Restore Input/Output Defaults" or "Restore System
Defaults->All"
Saved in instrument state.
On | Off
"Q Same as I" when On, otherwise none.
Prior to A.02.00
Input/Output, I/Q, Q Setup
:INPut:IQ:Q:DIFFerential OFF|ON|0|1
:INPut:IQ:Q:DIFFerential?
Put the Q channel in Differential Input mode
INP:IQ:Q:DIFF ON

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N9030aN9020aN9010aN9000aN9038a