Dependencies
Preset
State Saved
Min
Max
Readback
Backwards Compatibility SCPI
Initial S/W Revision
BTS
Sets an external attenuation value for BTS (Base Transceiver Station) tests.
Key Path
Remote Command
Example
Notes
The reference level limits are determined in part by the External Gain, Max
Mixer Level, RF Atten
This key is grayed out in modes that do not support MS.
This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output
Defaults" or "Restore System Defaults->All"
Saved in instrument state.
–100 dB
100 dB
MS, <Ext Gain value> dB
[:SENSe]:CORRection:MS[:RF]:LOSS <rel_ampl>
[:SENSe]:CORRection:MS[:RF]:LOSS?
Important notes regarding the alias commands:
A positive value of <rel_ampl> in the above command means a loss and a
negative value indicates a gain. So, for example, sending the command
a) CORR:MS:LOSS 10 dB will set the value on the softkey and the active
function to –10 dB since the softkey and the active function always show the
Gain.
The query CORR:MS:LOSS? returns 10 dB
The query CORR:MS:GAIN? returns –10 dB
b) CORR:MS:LOSS –10 dB will set the value on the softkey and the active
function to 10 dB since the softkey and the active function always show the
Gain
The query CORR:MS:LOSS? returns –10 dB
The query CORR:MS:GAIN? returns 10 dB
Prior to A.02.00
Input/Output, External Gain
[:SENSe]:CORRection:BTS[:RF]:GAIN <rel_ampl>
[:SENSe]:CORRection:BTS[:RF]:GAIN?
CORR:BTS:GAIN 10 sets the Ext Gain value to 10 dB
CORR:BTS:GAIN –10 sets the Ext Gain value to –10 dB (that is, a loss of 10
dB.)
Does not auto return.
Input/Output
433