Agilent Technologies X Series Manual page 450

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Input/Output
Scale X Axis
Matches the X Axis to the selected Correction, as well as possible. Sets the Start and Stop Frequency to
contain the minimum and maximum Frequency of the selected Correction. The range between Start
Frequency and Stop Frequency is 12.5% above the range between the minimum and maximum
Frequency, so that span exceeds this range by one graticule division on either side. If in zero-span, or
there is no data in the Ampcor table, or the frequency range represented by the table is zero, no action is
taken. Standard clipping rules apply if the value in the table is outside the allowable range for the X axis.
Key Path
Initial S/W Revision
Delete Correction
Deletes the correction values for this set. When this key is pressed a prompt is placed on the screen that
says "Please press Enter or OK key to delete correction. Press ESC or Cancel to close this dialog." The
deletion is only performed if you press OK or Enter.
Key Path
Remote Command
Example
Notes
Initial S/W Revision
Apply Corrections
Applies amplitude corrections which are marked as ON to the measured data. If this is set to OFF, then
no amplitude correction sets will be used, regardless of their individual on/off settings. If set to ON, the
corrections that are marked as ON (see
Key Path
Remote Command
Example
Preset
State Saved
Initial S/W Revision
450
Input/Output, Corrections, Edit
A.02.00
Input/Output, Corrections
[:SENSe]:CORRection:CSET[1]|2|3|4|5|6:DELete
CORR:CSET:DEL
CORR:CSET1:DEL
CORR:CSET4:DEL
Pressing this key when no corrections are present is accepted without error.
A.02.00
"Correction On/Off" on page
Input/Output, Corrections
[:SENSe]:CORRection:CSET:ALL[:STATe] ON|OFF|1|0
[:SENSe]:CORRection:CSET:ALL[:STATe]?
SENS:CORR:CSET:ALL OFF
This command makes sure that no amplitude corrections are applied,
regardless of their individual on/off settings.
Not affected by Preset. Set to OFF by Restore Input/Output Defaults
Saved in instrument state.
A.02.00
442) are used.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N9030aN9020aN9010aN9000aN9038a

Table of Contents