Agilent Technologies X Series Manual page 217

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

n
Results Returned
13
Returns the following comma-separated scalar results:
1. I/Q Magnitude and Phase Delta Results available (0 = not available, 1 = available).
2. Delta magnitude and phase trace (results 14 – 17) start frequency (0 when I/Q
3. Delta magnitude and phase trace (results 14 – 17) number of points (1 when I/Q
4. Delta magnitude and phase trace (results 14 – 17) frequency spacing between points (0
5. Current average count (1 when I/Q Magnitude and Phase Delta Results not available)
6. Frequency of the FFT trace (result 4) peak magnitude in Hz (0 when I/Q Magnitude
7. Delta magnitude at the FFT trace peak magnitude frequency in dB (–999 when I/Q
8. Delta phase at the FFT trace peak magnitude frequency in radians (–999 when I/Q
9. Delta phase at the FFT trace peak magnitude frequency in degrees (–999 when I/Q
10. Frequency of the averaged FFT trace (result 7) peak magnitude in Hz (0 when I/Q
11. Averaged delta magnitude at the averaged FFT trace peak magnitude frequency in dB
12. Averaged delta phase at the averaged FFT trace peak magnitude frequency in radians
13. Averaged delta phase at the averaged FFT trace peak magnitude frequency in degrees
14
Current delta magnitude trace in dB (–999 when I/Q Magnitude and Phase Delta Results
not available)
15
Current delta phase trace in radians (–999 when I/Q Magnitude and Phase Delta Results
not available)
16
Averaged delta magnitude trace in dB (–999 when I/Q Magnitude and Phase Delta Results
not available)
17
Averaged delta phase trace in radians (–999 when I/Q Magnitude and Phase Delta Results
not available)
Complex Spectrum Measurement Description
This measurement is FFT (Fast Fourier Transform) based. The FFT-specific parameters are located in the
Results are available when the last measurement was made with I/Q Magnitude and
Phase Delta Results enabled (SPEC:IQD:ENAB ON) and the setup was valid for
generating the results (invalid setup when input is I/Q, I/Q Path is I+jQ, and Center
Frequency is not 0 Hz)
Magnitude and Phase Delta Results not available)
Magnitude and Phase Delta Results not available)
when I/Q Magnitude and Phase Delta Results not available)
and Phase Delta Results not available)
Magnitude and Phase Delta Results not available)
Magnitude and Phase Delta Results not available)
Magnitude and Phase Delta Results not available or invalid setup)
Magnitude and Phase Delta Results not available)
(–999 when I/Q Magnitude and Phase Delta Results not available)
(–999 when I/Q Magnitude and Phase Delta Results not available)
(–999 when I/Q Magnitude and Phase Delta Results not available)
Complex Spectrum Measurement
217