Agilent Technologies X Series Manual page 478

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Input/Output
Notes
Initial S/W Revision
Back
Return to the prior step in the calibration procedure.
Key Path
Initial S/W Revision
Next
Perform the I port calibration.
Key Path
Remote Command
Example
Notes
State Saved
Initial S/W Revision
Exit
Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration
data.
Key Path
Notes
Initial S/W Revision
I-bar Port
The I-bar port calibration is performed with the probe body attached to the front panel's I-bar port and the
probe tip connected via an adapter to the Cal Out port. The I-bar probe calibration is only available for
passive probes with Differential On. The guided calibration will show a diagram of the required
connections.
478
Either a passive or an active probe adapter diagram will be shown, depending
on the type of probe attached.
Prior to A.02.00
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
Prior to A.02.00
Input/Output, I/Q, I Setup, I Probe, Calibrate
:CALibration:IQ:PROBe:I
CAL:IQ:PROB:I
The I port must be connected to the Cal Out port before issuing the SCPI
command.
The calibration data is saved as soon as the port is calibrated and will survive
power cycles. It is not reset by any preset or restore data commands.
No
Prior to A.02.00
Input/Output, I/Q, I Setup, I Probe, Calibrate
Using the Exit button will not restore the calibration data to the state prior to
entering the guided calibration. Once a port is calibrated the data is stored
immediately and the only way to change it is to redo the calibration step.
When the calibration may be left in an inconsistent state, a confirmation
dialog will be displayed (see
Prior to A.02.00
"Exit Confirmation" on page
482).