Agilent Technologies X Series Manual page 461

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

State Saved
Readback line
Backwards Compatibility Notes
Initial S/W Revision
More Information
The table below gives the range for each output.
Analog Out
Nominal Range
exc.
(10%
overrange)
Off
0 V
Screen
0 – 1 V open
Video
circuit
Log Video
0 – 1 V
terminated
Linear
0 – 1 V
Video
terminated
Demod
(varies with analyzer setting)
Audio
Auto
Selects the Auto state for the Analog Output menu. In this state, the Analog Output will automatically be
set to the most sensible setting for the current mode or measurement.
If you make a selection manually from the Analog Out menu, this selection will remain in force until you
change it (or re-select Auto), even if you go to a mode or measurement for which the selected output
does not apply.
Key Path
Remote Command
Example
Preset
Saved in Input/Output State
1-of-N selection [variable]
Prior to A.04.00, OFF was the default functionality except when in the Analog
Demod application or with Tune and Listen, in which case it was DAUDio,
and there was no selection menu. So for backwards compatibility, Auto
(:OUTP:ANAL:AUTO ON) will duplicate the prior behavior.
The DNWB and SANalyzer parameters, which were legal in PSA but perform
no function in the X-Series, are accepted without error.
A.04.00
Scale Factor
Notes
10%/division
8566 compatible
1/(192.66 dB/V)
dB referenced to mixer level, 1V out for –10
dBm at the mixer.
100%/V
Linear referenced to Ref Level, 1 V out for RF
envelope at the Ref Level.
Input/Output, Output Config, Analog Out
OUTPut:ANALog:AUTO OFF|ON|0|1
OUTPut:ANALog:AUTO?
OUTP:ANAL:AUTO ON
ON
Input/Output
461