Agilent Technologies X Series Manual page 466

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Input/Output
Readback Text
Initial S/W Revision
Modified at S/W Revision
I/Q Cal Out
The Baseband I/Q "Cal Out" port can be turned on with either a 1 kHz or a 250 kHz square wave. This
can be turned on independent of the input selection. A Preset will reset this to Off.
Key Path
Remote Command
Example
Couplings
Preset
State Saved
Range
Readback Text
Initial S/W Revision
Saved State
1 kHz Square Wave
Turns on the 1 kHz square wave signal at the Cal Out port. This choice is only available with option
BBA.
Key Path
Readback
Initial S/W Revision
250 kHz Square Wave
Turns on the 250 kHz square wave signal at the Cal Out port. This choice is only available with option
BBA.
Key Path
Readback
Initial S/W Revision
466
Demod Audio
Prior to A.02.00 (this was the default functionality, and there was no selection)
A.04.00
Input/Output, Output Config
:OUTPut:IQ:OUTPut IQ1|IQ250|OFF
:OUTPut:IQ:OUTPut?
OUTP:IQ:OUTP IQ1
An I/Q Cable Calibration or an I/Q Probe Calibration will change the state of
the Cal Out port as needed by the calibration routine. When the calibration is
finished the I/Q Cal Out is restored to the pre-calibration state.
Off
Saved in instrument state.
1 kHz Square Wave|250 kHz Square Wave|Off
1 kHz|250 kHz|Off
Prior to A.02.00
Saved in instrument state
Input/Output, Output Config, I/Q Cal Out
I/Q 1kHz
Prior to A.02.00
Input/Output, Output Config, I/Q Cal Out
I/Q 250kHz
Prior to A.02.00