Example
Couplings
Preset
State Saved
Backwards Compatibility SCPI
Initial S/W Revision
Trig Delay
Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger
criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT, but not in swept
spans.
Key Path
Remote Command
Example
Notes
Preset
State Saved
Min
Max
Default Unit
TRIG:EXT1:SLOP NEG
This same slope is used in the Ext1 selection for the trigger source in the
Trigger menu and for the period timer sync source (in the Trigger menu and in
the Gate Source menu).
POSitive
Saved in instrument state
For backward compatibility, EXTernal should also work.
Also, there are SLOPe backward compatibility commands described in the
Video section
"Trig Slope " on page 578
Prior to A.02.00
Trigger, External 1
:TRIGger[:SEQuence]:EXTernal1:DELay <time>
:TRIGger[:SEQuence]:EXTernal1:DELay?
:TRIGger[:SEQuence]:EXTernal1:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:EXTernal1:DELay:STATe?
TRIG:EXT1:DEL:STAT ON
TRIG:EXT1:DEL 100 ms
Video trigger delay may be set to negative values, in time domain, FFT and
even swept. It makes intuitive sense in time domain and works well in FFT
mode where the bandwidth of the filter before the video trigger is about 1.25
span. In swept spans, negative settings of Trig Delay are treated as a zero
setting within the internal hardware and the advisory message "Neg. Trig
Delay unavailable in Swept Mode, zero delay used." is generated when such a
delay is set.
Off, 1.000 us
Saved in instrument state
–150 ms
+500 ms
s
:TRIGger[:SEQuence]:SLOPe