Input/Output
State Saved
Initial S/W Revision
Exit
Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired calibration
data.
Key Path
Notes
Initial S/W Revision
Show Adapter Screen
When one of the Probe Calibration Show Adapter buttons is pressed, a diagram of the probe with its
adapter will be shown. Depending on the type of probe attached, either the Passive Probe Adapter or the
Active Probe Adapter diagram will be shown.
I/Q Probe Calibration Time (Remote Command Only)
Return the last date and time that the I/Q Probe Calibration was performed for a specific port. This is a
remote query command only.
Remote Command
Example
Notes
Initial S/W Revision
Exit Confirmation
When Exit is pressed during one of the calibration routines, the calibration may be in an inconsistent
state with some of the ports having newly measured calibration data and others with old data. If this is
the case, a dialog box will appear to confirm that the user really wants to exit. A "Yes" answer will exit
the calibration procedure, leaving potentially inconsistent calibration data in place. A "No" answer will
return to the calibration procedure.
482
No
Prior to A.02.00
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
Using the Exit button will not restore the calibration data to the state prior to
entering the guided calibration. Once a port is calibrated the data is stored
immediately and the only way to change it is to redo the calibration step.
When the calibration may be left in an inconsistent state, a confirmation
dialog will be displayed (see
Prior to A.02.00
:CALibration:IQ:PROBe:I|IBAR|Q|QBAR:TIME?
:CAL:IQ:PROB:I:TIME?
This returns 6 integer values: year, month, day, hour, minute, second. When no
calibration has been performed, all values will be 0. The value is specific to
both the port and probe, so the value will change as probes are connected or
disconnected.
A.02.00
"Exit Confirmation" on page
482).